PreviousNext

Default / Reset / Save

Understanding and controlling your Imager's active, saved, and default settings is critical to its successful operation.

 
Function
Serial Cmd
ESP (first, right click in any menu tree)1
EZ Button
Resets
(not saved
for power-on)
Reset
<A>
Save to Reader,
Send No Save
No
Reset and recall Microscan defaults
<Ard>
No
No
Reset and recall power-on parameters
<Arp>
No
No
Reset and recall customer default parameters
<Arc>
No
No
Saved for Power-on
Save current settings for power-on
<Z>
Save to Reader,
Send and Save
No
Save current settings as customer default parameters for power-on
<Zc>
Save to Reader,
Send and Save Customer Defaults for Power-On
2
No
Recall Microscan default parameters and save for power-on
<Zrd>
No
No
Recall customer default parameters and save for power-on
<Zrc>
No
Press and hold while powering on Imager

Resets

Resets ("A" commands) affect only the current settings (active memory) and are not saved for power-on.

Saved for Power-on

Power-on parameters ("Z" commands) are saved to NOVRAM and recalled and loaded into current parameters when power is cycled to the Imager or the <Arp> command is sent.

Defaults

Defaults are Microscan firmware settings or saved customer settings that can be recalled, either by software or hardware reset.