PreviousNext

Error Codes

In the event of an error being detected, the beeper beeps five times and an error code (in binary form) representing the test that failed flashes on the LEDs. The LEDs continue to flash until the <Escape> character is sent on the serial port, at which point an error message is logged to the serial port and the Diagnostic Monitor is launched. The serial port terminal server must be set to the following parameters: 115200, 8, N, 1.

Table 2-5 describes the Mode/Status LED power-on sequence and error codes.

TABLE 2-5. Mode/Status LED Power-On Sequence and Error Codes

Mode

Fail

Pass

Trig

Test Performed

     
·
Data Line Test
   
·
 
Address Line Test
   
·
·
Bootloader CRC Check
 
·
   
Kernel CRC Check
 
·
 
·
RS-232 Internal Loopback
 
·
·
 
FPGA Load Test
 
·
·
·
FPGA PCI Config Test
·
     
FPGA Video Buffer Test
·
   
·
FPGA DMA Transfer Test
·
 
·
 
FPGA Expose Done Interrupt Test
·
 
·
·
FPGA Read Done Interrupt Test


PreviousNext